首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Supported liquid extraction in combination with LC‐MS/MS for high‐throughput quantitative analysis of hydrocortisone in mouse serum
Authors:Shari Wu  Wenkui Li  Tapan Mujamdar  Tom Smith  Matthew Bryant  Francis L S Tse
Institution:Department of Drug Metabolism and Pharmacokinetics, Novartis Institutes for Biomedical Research, One Health Plaza, East Hanover, NJ 07936, USA
Abstract:A high‐throughput LC–MS/MS bioanalytical method was developed and validated for the determination of hydrocortisone in mouse serum via supported liquid extraction (SLE) in a 96‐well plate format. Although sample extracts from SLE result in similar matrix effects compared with conventional liquid–liquid extraction (LLE), greater analyte extraction recovery and much higher analysis throughput for the quantitative analysis of hydrocortisone in mouse serum were obtained. The current LC‐MS/MS method was validated for a concentration range of 2.00–2000 ng/mL for hydrocortisone using a 0.100 mL volume of mouse serum. The intra‐ and inter‐day precision and accuracy of the quality control samples at low, medium and high concentration levels showed ≤12.9% CV and ?3.4–6.2% bias for the analyte in mouse serum. Copyright © 2009 John Wiley & Sons, Ltd.
Keywords:supported liquid extraction  hydrocortisone  liquid–  liquid extraction  LC‐MS/MS  mouse serum
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号