Efforts to improve detection sensitivity for capillary electrophoresis coupled to atmospheric pressure photoionization mass spectrometry
Authors:Jakob Axén  David Malmström  Bengt‐Olof Axelsson  Patrik Petersson  Per J R Sjöberg
Institution:1. Department of Physical and Analytical Chemistry, Analytical Chemistry, Uppsala University, Box 599, SE‐751 24 Uppsala, Sweden;2. AstraZeneca R&D Lund, Analytical Development, SE‐221 87 Lund, Sweden
Abstract:Electrospray ionization performs best with volatile buffers. However, generally the best separation performance for capillary electrophoresis (CE) is achieved with non‐volatile buffers. Hyphenation of CE with mass spectrometry (MS) utilizing atmospheric pressure photoionization (APPI) enables use of a wider range of separation buffers without compromising detection sensitivity. As APPI is considered to be mass flow sensitive, the use of a larger inner diameter separation capillary (75 µm) allows larger volumes to be injected, without decreased separation performance, thus providing improved sensitivity (approx. a factor of 10), compared to the use of a 25 µm capillary. However, nebulizing gas flow and position of capillary tip in the sprayer have to be carefully optimized to prevent excessive band broadening. Further improvement in sensitivity (approx. a factor of 2) was obtained by decreasing the distance between the sprayer and ionization region, indicating that a specially designed CE/APPI‐MS interface for low flow rates will be favourable. Copyright © 2010 John Wiley & Sons, Ltd.