首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Optimization of the ICP working conditions for the analysis of waste waters and industrial solutions
Authors:A Delijska and Z Zadgorska
Institution:(1) Research and Design Institute of Mining and Ore Dressing ldquoNIPRORUDArdquo, Stambolijski 205, 1309 BG-Sofia, Bulgaria;(2) Institute of General and Inorganic Chemistry, Bulgarian Academy of Sciences, 1040 BG-Sofia, Bulgaria
Abstract:Summary In order to optimize the working parameters of ICP analysis of waste waters and industrial solutions the spectral interferences and the effect of the carrier gas flow were studied. Special attention was paid to the influence of high contents of salt and H2SO4 in some samples. The use of a peristaltic pump, dilution or internal standardization enable reliable analytical results to be obtained. A direct ICP method for determination of Al, Ca, Cd, Cu, Fe, Mg, Mn, Ni, Pb and Zn was worked out. Relative standard deviations were in the range of 0.5–5%.
Optimierung der ICP-Arbeitsbedingungen für die Analyse von Abwasser und industriellen Lösungen
Zusammenfassung Zur Optimierung der ICP-Parameter wurden die spektralen Störungen sowie der Einfluß des Trägergasstromes untersucht. Der Wirkung hoher Salz- und Schwefelsäuregehalte wurde besondere Aufmerksamkeit gewidmet. Mit Hilfe einer peristaltischen Pumpe sowie durch entsprechende Verdünnung und Anwendung des Verfahrens des inneren Standards konnten zuverlässige Resultate erzielt werden. Eine Arbeitsmethode wurde entwickelt für die Bestimmung von Al, Ca, Cd, Cu, Fe, Mg, Mn, Ni, Pb und Zn. Die Standardabweichungen lagen im Bereich von 0,5–5%.