首页 | 本学科首页   官方微博 | 高级检索  
     


Investigation of chromium(III) acetylacetonate as a calibration reference material for atomic absorption spectroscopy
Authors:Oleg Yurchenko  Konstantin Belikov  Nikolay Shevtsov
Affiliation:(1) V.N. Karazin Kharkov National University, Kharkov, Ukraine;(2) State Scientific Institution “Institute for Single Crystals” of the National Academy of Sciences of Ukraine, Kharkov, Ukraine
Abstract:It is shown that chromium(III) acetylacetonate meets the requirements of calibration reference materials: it is homogeneous, soluble in water and organic solvents, and its solutions are stable. Chromium(III) acetylacetonate has been successfully used as a calibration reference material in the analysis of chromium in iron-base alloys. The procedure involves chromium acetylacetonate extraction using methyl isobutyl ketone followed by direct atomic absorption analysis of the organic extract. The detection limit of the procedure is 10 ng g−1. Correspondence: Konstantin Belikov, State Scientific Institution “Institute for Single Crystals” of the National Academy of Sciences of Ukraine, Kharkov, Ukraine
Keywords:: Atomic absorption spectrometry   reference material   chromium   acetylacetonate   iron-base alloys
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号