微量进样/ICP-MS体系中的基体效应研究
引用本文:康建珍,段太成,郭鹏然,王畅,陈杭亭,曾宪津.微量进样/ICP-MS体系中的基体效应研究[J].高等学校化学学报,2004,25(2):252-255.
作者姓名:康建珍  段太成  郭鹏然  王畅  陈杭亭  曾宪津
作者单位:中国科学院长春应用化学研究所电分析化学国家重点实验室, 国家电化学和光谱研究分析中心, 长春 130022
基金项目:国家自然科学基金 (批准号 :2 0 0 75 0 2 9)资助
摘    要:考察了自行组装的高效微量进样系统在22μL/min低提升量下电感耦合等离子体质谱的基体效应.质量数和电离电位不同的基体元素质量浓度为2g/L时,低质量数分析元素受到一定的干扰,而对于高质量数分析元素,其信号几乎不受基体元素干扰,归一化信号值接近1.对于体积分数为5%的有机基体样品溶液,此微量高效雾化系统测得的归一化信号值多接近0.5.实际样品中常见的基体元素K,Na,Ca和Mg质量浓度低于500mg/L,以及乙醇和乙酸体积分数小于1%时,微量进样系统均不产生显著干扰.

关 键 词:电感耦合等离子体质谱  高效微量进样系统  基体效应  
文章编号:0251-0790(2004)02-0252-04
收稿时间:2002-09-16

Matrix Effects of Micro-sampling System for Inductively Coupled Plasma Mass Spectrometry
KANG Jian-Zhen,DUAN Tai-Cheng,GUO Peng-Ran WANG Chang,CHEN Hang-Ting ,ZENG Xian-Jin.Matrix Effects of Micro-sampling System for Inductively Coupled Plasma Mass Spectrometry[J].Chemical Research In Chinese Universities,2004,25(2):252-255.
Authors:KANG Jian-Zhen  DUAN Tai-Cheng  GUO Peng-Ran WANG Chang  CHEN Hang-Ting  ZENG Xian-Jin
Institution:State Key Laboratory of Electroanalytical Chemistry, National Analytical Research Center of Electrochemistry & Spectroscopy, Changchun Institute of Applied Chemistry, Chinese Academy of Sciences, Changchun 130022, China
Abstract:Matrix effects of inductively coupled plasma mass spectrometry(ICP-MS) with the self-installed micro-sampling system at 22 μL/min of solution uptake rate were studied. In the presence of 2 g/L matrix elements with different masses or ionization potentials, the signal intensity of analyzed elements was measured. For light analyzed elements, the interferences were observed; while for high mass analyte elements, the normalized intensity was close to 1. In the presence of 5% organic matrix, the normalized intensity was near to 0 5. When common elements such as K, Na, Ca, Mg at 500 mg/L or 1% organic solvents such as ethanol and acetic acid existed in the solution, the signals of analyzed elements were almost not affected. The results show that the micro-sampling system has a good capability of anti-interference to matrixes.
Keywords:Inductively coupled plasma mass spectrometry  High effective micro-sampling system  Matrix effects  