首页 | 本学科首页   官方微博 | 高级检索  
     检索      


The amperometric detection of thyroid hormones following reverse-phase high-performance liquid chromatography
Authors:Bradford R Hepler  Stephen G Weber  William C Purdy
Institution:Department of Chemistry, McGill University, 801 Sherbrooke St. W., Montreal, Quebec, H3A 2K6 Canada
Abstract:The principle of an assay of the major thyroid hormones by an electrochemical technique is demonstrated. The separation of 3,3',5-triiodothyronine, 3,3',5'-triiodothyronine, and thyroxine, by reverse-phase high-performance liquid chromatography is followed by their electrochemical oxidation in a thin-layer electrochemical detection cell with a low-temperature isotropic carbon working electrode. The limits of detection found were in the subnanogram range with linear response in the ranges 0–125 ng for T3 and 0–500 ng for T4. The approach makes the simultaneous assay of total serum thyroid hormones feasible.
Keywords:
本文献已被 ScienceDirect 等数据库收录!