首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Mucin/carbopol matrix to immobilize oxalate oxidase in a urine oxalate amperometric biosensor
Authors:RH Capra  PM Vadgama
Institution:a INFIQC, Departamento de Fisicoquímica, Facultad de Ciencias Químicas, Universidad Nacional de Córdoba, Argentina
b IMBIV, Departamento de Química Orgánica, Facultad de Ciencias Químicas, Universidad Nacional de Córdoba, Argentina
c IRC in Biomedical Materials, Queen Mary, University of London, England
d Hospital Privado Centro Médico de Córdoba, Argentina
Abstract:An enzymatic amperometric electrode with extended analytical range and improved stability for oxalate determination has been developed. Glutarlaldehyde/mucin/carbopol matrix was used for the crosslinking of the enzyme between polymeric membranes to form a classical laminate construction (sandwich) and compared with the glutaraldehyde/mucin/enzyme and glutaraldehyde/albumin/enzyme.The use of a sulphonated membrane as internal membrane allowed rejection of the most important electrooxidable urine interferents. The recovery assays were highly satisfactory. The wide linear response in the range 2-400 μM after 1/10 urine dilution (corresponding to 20-4000 μM) made it suitable for clinical range. High correlation with the standard spectrophotometric method was obtained (r2 = 0.98, y = 0.89x, n = 25).
Keywords:Mucin  Carbopol  Oxalate biosensor  Amperometric biosensor  Oxalate oxidase
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号