Oscillatory disturbance in force calibration of optical tweezers
引用本文:刘春香,郭红莲,降雨强,李兆霖,程丙英,张道中. Oscillatory disturbance in force calibration of optical tweezers[J]. 中国物理, 2005, 14(4): 729-733
作者姓名:刘春香  郭红莲  降雨强  李兆霖  程丙英  张道中
作者单位:Optical Physics Laboratory, Institute of Physics and Centre for Condensed Matter Physics, Chinese Academy of Sciences, Beijing,100080, China;Optical Physics Laboratory, Institute of Physics and Centre for Condensed Matter Physics, Chinese Academy of Sciences, Beijing,100080, China;State Key Laboratory of Quantum Optics and Quantum Optics Devices, Department of Electronics and Information Technology,Shanxi University, Taiyuan,030006, China;Optical Physics Laboratory, Institute of Physics and Centre for Condensed Matter Physics, Chinese Academy of Sciences, Beijing,100080, China;Optical Physics Laboratory, Institute of Physics and Centre for Condensed Matter Physics, Chinese Academy of Sciences, Beijing,100080, China;Optical Physics Laboratory, Institute of Physics and Centre for Condensed Matter Physics, Chinese Academy of Sciences, Beijing,100080, China
基金项目:Project supported by the National Key Basic Research Special Foundation of China (Grant No 2002CB713805) and the National Natural Science Foundation of China (Nos 10174097 and 10374112)
摘    要:In the calibration of the optical trap stiffness, it is found that there appears an attenuating oscillation as an oscillatory disturbance added to the trapped bead movement, when the scanner is driven by a triangular wave input.An equivalent oscillator model is put forward to explain the mechanism of the oscillatory disturbance. Both the measurements and calculations show that the attenuating oscillation comes from the oscillation of the scanner and the triangular wave drive causes this additional oscillation of the scanner. Furthermore, the analysis indicates that the oscillatory disturbance will become stronger, when the stiffness of the trap increases or the natural frequency of the scanner decreases. We adopt another driving way, i.e. a sinusoidal wave input is used instead of the triangular wave input. Our experiment has verified that in this case the oscillatory disturbance is eliminated completely.

关 键 词:光学镊子 力度校准 振动俘获 硬度俘获 流体力学
收稿时间:2004-03-31

Oscillatory disturbance in force calibration of optical tweezers
Liu Chun-Xiang,Guo Hong-Lian,Jiang Yu-Qiang,Li Zhao-Lin,Cheng Bing-Ying and Zhang Dao-Zhong. Oscillatory disturbance in force calibration of optical tweezers[J]. Chinese Physics, 2005, 14(4): 729-733
Authors:Liu Chun-Xiang  Guo Hong-Lian  Jiang Yu-Qiang  Li Zhao-Lin  Cheng Bing-Ying  Zhang Dao-Zhong
Affiliation:Optical Physics Laboratory, Institute of Physics and Centre for Condensed Matter Physics, Chinese Academy of Sciences, Beijing,100080, China; State Key Laboratory of Quantum Optics and Quantum Optics Devices, Department of Electronics and Information Technology,Shanxi University, Taiyuan,030006, China
Abstract:In the calibration of the optical trap stiffness, it is found that there appears an attenuating oscillation as an oscillatory disturbance added tothe trapped bead movement, when the scanner is driven by a triangular waveinput. An equivalent oscillator model is put forward to explain themechanism of the oscillatory disturbance. Both the measurements andcalculations show that the attenuating oscillation comes from theoscillation of the scanner and the triangular wave drive causes thisadditional oscillation of the scanner. Furthermore, the analysis indicatesthat the oscillatory disturbance will become stronger, when the stiffness ofthe trap increases or the natural frequency of the scanner decreases. Weadopt another driving way, i.e. a sinusoidal wave input is used instead ofthe triangular wave input. Our experiment has verified that in this case theoscillatory disturbance is eliminated completely.
Keywords:optical tweezers   force calibration   oscillatory disturbance