Lineare Silizium-Photodioden-Matrizen mit parallelem Datenausgang als Strahlungsempfänger in der optischen Emissionsspektroskopie—I. Messanordnung und Analysen mit Funkenstrahlungsquelle |
| |
Authors: | H Bubert W-D Hagenah K Laqua |
| |
Institution: | Institut für Spektrochemie und angewandte Spektroskopie, Bunsen-Kirchhoff-Straße 11, D-4600 Dortmund 1, Bundesrepublik Deutschland |
| |
Abstract: | A linear silicon photodiode array has been used which consists of 5 photodiodes each having a separate output. Because of the dimensions of a single photodiode, i.e. 3mm long and 20 μm wide, they are capable of replacing the exit slits of a spectrometer. The read out system having 3 channels is described in detail. In each channel a FET operational amplifier with a low input bias current served as a current-to-voltage converter.The signal-to-noise ratios of systems using the photodiode array and of a photomultiplier (EMI 9789 QB) respectively are compared. The dynamic range of the photodiode is linear and extends over 5 decades referred to the noise level. As an analytical example copper was determined in aluminium alloys using medium voltage spark excitation in an argon atmosphere. The limit of detection was c = 4.8 × 10?3%, the relative standard deviation as a measure of the precision was srel = 0.014 for concentrations above 0.08%.In part II the application of recently produced photodiode arrays on the analysis of powders with a glow discharge will be described. |
| |