首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Lineare Silizium-Photodioden-Matrizen mit parallelem Datenausgang als Strahlungsempfänger in der optischen Emissionsspektroskopie—I. Messanordnung und Analysen mit Funkenstrahlungsquelle
Authors:H Bubert  W-D Hagenah  K Laqua
Institution:Institut für Spektrochemie und angewandte Spektroskopie, Bunsen-Kirchhoff-Straße 11, D-4600 Dortmund 1, Bundesrepublik Deutschland
Abstract:A linear silicon photodiode array has been used which consists of 5 photodiodes each having a separate output. Because of the dimensions of a single photodiode, i.e. 3mm long and 20 μm wide, they are capable of replacing the exit slits of a spectrometer. The read out system having 3 channels is described in detail. In each channel a FET operational amplifier with a low input bias current served as a current-to-voltage converter.The signal-to-noise ratios of systems using the photodiode array and of a photomultiplier (EMI 9789 QB) respectively are compared. The dynamic range of the photodiode is linear and extends over 5 12 decades referred to the noise level. As an analytical example copper was determined in aluminium alloys using medium voltage spark excitation in an argon atmosphere. The limit of detection was c = 4.8 × 10?3%, the relative standard deviation as a measure of the precision was srel = 0.014 for concentrations above 0.08%.In part II the application of recently produced photodiode arrays on the analysis of powders with a glow discharge will be described.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏