首页 | 本学科首页   官方微博 | 高级检索  
     

凝胶渗透色谱净化-高效液相色谱法测定油脂食品中的邻苯二甲酸酯类增塑剂
引用本文:Zhang C,Wang H,Zhang X,Ma Z,Deng W,Hu K,Ding M. 凝胶渗透色谱净化-高效液相色谱法测定油脂食品中的邻苯二甲酸酯类增塑剂[J]. 色谱, 2011, 29(12): 1236-1239. DOI: 10.3724/SP.J.1123.2011.01236
作者姓名:Zhang C  Wang H  Zhang X  Ma Z  Deng W  Hu K  Ding M
作者单位:1. 北京莱伯泰科有限公司, 北京 101312; 2. 清华大学化学系, 北京 100084
基金项目:国家自然科学基金项目(No. 21075074).
摘    要:建立了油脂食品(方便面、油炸糕点、沙琪玛、食用油等)中5种主要邻苯二甲酸酯类增塑剂的凝胶渗透色谱(GPC)净化-高效液相色谱(HPLC)分析方法。食品样品用石油醚超声提取,经GPC净化后,采用反相HPLC进行分析。所用的分离柱为Labtech C18(250 mm×4.6 mm, 5 μm),以乙腈和水为流动相,梯度洗脱。方法的相关系数皆在0.997以上,目标物的检出限(信噪比为3计)为3.25~13.4 μg/L。在50 mg/L添加水平时,目标物的加标回收率为70.4%~113.6%,相对标准偏差为0.3%~5.8%(n=3)。该方法简便、快捷、实用,可用于油脂食品中邻苯二甲酸酯类增塑剂的分析测定。

关 键 词:凝胶渗透色谱  高效液相色谱  邻苯二甲酸酯  增塑剂  油脂食品
收稿时间:2011-08-17

Determination of phthalate plasticizers in foods by high performance liquid chromatography with gel permeation chromatographic clean-up
Zhang Chunyu,Wang Hui,Zhang Xiaohui,Ma Zhongqiang,Deng Wanmei,Hu Ke,Ding Mingyu. Determination of phthalate plasticizers in foods by high performance liquid chromatography with gel permeation chromatographic clean-up[J]. Chinese journal of chromatography, 2011, 29(12): 1236-1239. DOI: 10.3724/SP.J.1123.2011.01236
Authors:Zhang Chunyu  Wang Hui  Zhang Xiaohui  Ma Zhongqiang  Deng Wanmei  Hu Ke  Ding Mingyu
Affiliation:1. Beijing Labtech Group, Beijing 101312, China; 2. Department of Chemistry, Tsinghua University, Beijing 100084, China
Abstract:A method of gel permeation chromatography-high performance liquid chromatography (GPC-HPLC) was established for the simultaneous determination of 5 main phthalate plasticizers in foods (edible oil, instant noodles, fried pastries, Saqima, etc.). The samples were extracted with petroleum ether in an ultrasonator, purified by a GPC column, and analyzed by HPLC. The chromatographic separation was achieved on a Labtech-C18 column (250 mm x 4.6 mm, 5 microm) using acetonitrile and water mixture as the mobile phases in a gradient elution mode. The developed method exhibited a linear correlation coefficient of more than 0.997 and the detection limits of 3.25 - 13.4 microg/L. The spike recoveries were between 70.4% and 113.6% with the relative standard deviations (RSDs, n = 3) of 0.3% - 5.8% at the spiked level of 50 mg/L. This method is simple, rapid and practical, and can be used for the simultaneous determination of PAEs in grease food samples.
Keywords:gel permeation chromatography (GPC)   high performance liquid chromatography (HPLC)   phthalates   plasticizers   grease food
本文献已被 PubMed 等数据库收录!
点击此处可从《色谱》浏览原始摘要信息
点击此处可从《色谱》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号