A new approach for fluorine determination by solid sampling graphite furnace molecular absorption spectrometry |
| |
Authors: | Éder Lisandro de Moraes Flores Juliano Smanioto Barin Érico Marlon de Moraes Flores Valderi Luiz Dressler |
| |
Institution: | 1. Universidade Federal de Santa Maria, Departamento de Química, 97105-900, Santa Maria, RS, Brazil;2. Universidade Estadual do Estado do Rio Grande do Sul, 97573-011, Santana do Livramento, RS, Brazil;3. Universidade Regional Integrada do Alto Uruguai, 98400-000, Frederico Westphalen, RS, Brazil |
| |
Abstract: | This work deals with the determination of fluorine by solid sampling graphite furnace molecular absorption spectrometry. The molecular absorbance of aluminum monofluoride (AlF), which is produced in the vapor phase in the presence of Al3+, is measured at 227.5 nm, a non-resonant platinum line. A conventional graphite furnace program has been used with pyrolysis and vaporization temperatures of 800 and 2300 °C, respectively. Solutions of Ba2+ and Al3+ have been used to avoid fluorine losses during the pyrolysis stage and to produce AlF in the vaporization stage, respectively. Certified coal and alumina samples were analyzed using aqueous standards for calibration. The agreement between the found concentration and the certified value, or the value obtained by another method ranged from 92 to 105%, with a relative standard deviation less than 8.5%. The limit of detection and the characteristic mass was 0.17 μg g− 1 and 205 pg, respectively. |
| |
Keywords: | Fluorine Solid sampling Graphite furnace molecular absorption spectrometry Coal Alumina |
本文献已被 ScienceDirect 等数据库收录! |
|