A Si(Li) detector system for conversion coefficient measurements and application to the decays of51Cr and87Y
Authors:H K Carter  J H Hamilton
Institution:1. Physics Department, Vanderbilt University, Nashville, Tennessee, USA
Abstract:An electron detector system with a 3 mm × 100 mm2 Si(Li) detector with a cooled FET and vacuum air lock for changing sources was constructed. The absolute efficiency was measured as a function of energy for use in the measurement ofK conversion coefficients. With the Si(Li) and a Ge(Li) detector, theK conversion coefficient of the 320 keV transition in51V was measured to be 0.00156 ± 0.00008 where theK conversion coefficients of the 661 and 279 keV transitions in137Ba and203Tl were used as standards. From absolute detection efficiency measurements, theK conversion coefficients of the 388 and 483 keV transitions in87Sr were measured to be 0.16 ± 0.02 and 0.0024 ± 0.0003, respectively.