首页 | 本学科首页   官方微博 | 高级检索  
     


Use of CRM's as mutual calibrating materials and control of syntheti multielement standards as used in INAA
Authors:M. Rossbach  M. Stoeppler
Affiliation:(1) Institute of Applied Physical Chemistry, ICH-4, Nuclear Research Center KFA, Jülich, Germany
Abstract:The comparability of analytical results from different laboratories requires accurately known concentrations in the applied standards. Dilution effects of different multielement synthetic standard solutions have been studied by measuring 10–12 different concentrations of the same solution. Peak area comparison of four Certified Reference Materials (CRM's) using one value for the evaluation of the other three repetitively led to the intercomparison (degree of compatibility) of the certified values. The idea of the preparation of each laboratory's ldquosecondary reference standardrdquo by comparison of synthetic multielement standards with as many CRM's as practically feasible is advocated to improve the reliability of analytical results.
Keywords:
本文献已被 SpringerLink 等数据库收录!