首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Automated on-line separation preconcentration system for palladium determination by graphite furnace atomic absorption spectrometry and its application to palladium determination in environmental and food samples
Authors:Rojas F Sánchez  Ojeda C Bosch  Pavón J M Cano
Institution:

aDepartment of Analytical Chemistry, Faculty of Sciences, University of Málaga, E-29071 Málaga, Spain

Abstract:A flow injection (FI) system was used to develop an efficient on-line sorbent extraction preconcentration system for palladium by graphite furnace atomic absorption spectrometry (GFAAS). The investigated metal was preconcentrated on a microcolumn packed with 1,5-bis(di-2-pyridyl)methylene thiocarbohydrazide immobilized on silica gel (DPTH-gel). The palladium is eluted with 40 μl of HCl 4 M and directly introduced into the graphite furnace. The detection limit for palladium under the optimum conditions was 0.4 ng ml?1. This procedure was employed to determine palladium in different samples.
Keywords:Palladium  Graphite furnace atomic absorption spectrometry  Preconcentration  Flow injection
本文献已被 ScienceDirect PubMed 等数据库收录!