首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Stability indicating method for famotidine in pharmaceuticals using porous graphitic carbon column
Authors:Helali Nawel  Monser Lotfi
Institution:Institut National des Sciences Appliquées et de Technologie, Centre Urbain Nord, Tunis Cedex, Tunisia.
Abstract:A simple, sensitive and rapid HPLC method was developed and validated for the simultaneous determination of famotidine (FMT) and related impurities in pharmaceuticals. Chromatographic separation was accomplished within 10 min on a porous graphitic carbon (PGC) column using 50:50 v/v ACN-water containing 0.5% pentane sulphonic acid (PSA) as the mobile phase. Separation was achieved with a flow rate of 1 mL/min and a detection wavelength of 265 nm. The calibration curves were linear over a concentration range of 1.5-100 microg/mL. The intra- and interday RSDs (n = 5) for the retention times and peak area were all less than 2%. The method was sensitive with an LOD (S/N = 3) of 0.1 microg/mL for FMT, imp. C and 0.05 microg/mL for imp. 2, A and D. All recoveries were greater than 98%. The method was demonstrated to be precise, accurate and specific with no interference from the tablet ingredients and separation of the drug peak from the peaks of the degradation products (oxidative degradation and acid and base degradation). The results indicated that the proposed method could be used for the determination of FMT in commercial dosage forms and as a stability-indicating assay.
Keywords:Famotidine  HPLC  Porous graphitic carbon column  Related impurities  Tablets
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏