首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A low-leakage sample plug injection scheme for crossform microfluidic capillary electrophoresis devices incorporating a restricted cross-channel intersection
Authors:Chang Chin-Lung  Hou Hui-Hsiung  Fu Lung-Ming  Tsai Chien-Hsiung
Institution:Department of Vehicle Engineering, National Pingtung University of Science and Technology, Pingtung, Taiwan.
Abstract:This study develops a crossform CE microfluidic device in which a single-circular barrier or a double-circular barrier is introduced at the cross-channel intersection. Utilizing a conventional crossform injection scheme, it is shown that these barriers reduce sample leakage and deliver a compact sample band into the separation channel, thereby ensuring an enhanced detection performance. A series of numerical and experimental investigations are performed to investigate the effects of the barrier type and the barrier ratio on the flow streamlines within the microchannel and to clarify their respective effects on the sample leakage ratio and sample plug variance during the injection process. The results indicate that a single-circular barrier injector with a barrier ratio greater than 20% and a double-circular barrier injector with a barrier ratio greater than 40% minimize the sample leakage ratio and produce a compact sample plug. As a result, both injectors have an excellent potential for use in high-quality, high-throughput chemical analysis procedures and in many other applications throughout the micro-total analysis systems field.
Keywords:Capillary electrophoresis microfluidic device  Circular barrier  Sample leakage ratio  Sample plug variance
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏