Development of a dipstick immunoassay for quantitative determination of 2,4-dichlorophenoxyacetic acid in water, fruit and urine samples |
| |
Authors: | N V Cuong T T Bachmann and R D Schmid |
| |
Institution: | (1) National Institute of Biotechnology, Hoang Quoc Viet Street, Cau Giay, Hanoi, Vietnam, VN;(2) Institut für Technische Biochemie, Universit?t Stuttgart, Allmandring 31, D-70569 Stuttgart, Germany, DE |
| |
Abstract: | A sensitive dipstick assay for 2,4-dichlorophenoxyacetic acid (2,4-D) detection was developed. The assay was based on the
competitive reaction of 2,4-D and enzyme tracer with monoclonal antibodies immobilised on an Ultrabind? membrane. The binding of enzyme tracer on the test strip was determined by a simple, portable reflectometer as remission
at 657 nm. Using this technique, 2,4-D could be detected in a concentration range of 0.5 μg/L to 100 μg/L. The center point
of the 2,4-D test was found at a concentration of 6 μg/L. Cross-reactivity with 2,4,5-trichlorophenoxyacetic acid (2,4,5-T)
as determined by this dipstick assay was 2.5% and 3% by standard ELISA technique using microtiter plates. The assay was applied
in the detection of 2,4-D in real water samples, and sensitivity was comparable to spiked water samples. If combined with
an effective extraction procedure that results in recovery rates of 90%, the dipstick assay can be used to monitor human exposure
to 2,4-D from contamination in water, from oranges and in testing urine samples.