首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Rapid and accurate determination of89/90Sr in radioactive samples by Cerenkov counting
Authors:Tser-Min Chang  Shao-Chou Chen  Jwn-Yung King  Shieh-Jun Wang
Institution:(1) Chemical Analysis Division, Institute of Nuclear Energy Research, P.O. Box 3-19, 325 Lungtan, Taiwan (ROC)
Abstract:An improved and rapid method for determination of90Sr via its daughter nuclide,90Y, in aqueous samples from the low-level radioactive wastes by Cerenkov counting was established. This technique is applicable to beta particles maximum energies greater than 0.263 MeV in aqueous solution. A comparison of90Sr determination by Cerenkov counting and standard wet chemical separation techniques indicates a high degree of correlation and excellent agreement. For a 20 ml aqueous sample following the prescribed experimental conditions and a 60 min counting interval, the detection limit was 0.20 dpm/ml (0.10 pCi/ml), and the relative deviation is less then 5%.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号