An all-solid-state monohydrogen phosphate sensor based on a macrocyclic ionophore |
| |
Authors: | Pankaj Kumar |
| |
Institution: | Department of Chemistry and Institute of BioPhysio Sensor Technology (IBST), Pusan National University, Jangjun-Dong, Busan, 609-735, South Korea |
| |
Abstract: | An all-solid-state electrode, containing a synthesized chiral A2B2 macrocyclic compound namely (4R,5R,15R,16R)-4,5,15,16-tetraphenyl-3,6,14,17-tetraazatricyclo 13.3.1.18,12] tetracosa-1(23),8,10,12(24)19,21-hexaene-2,7,13,18-tetrone as an ionophore in polyvinyl chloride (PVC)/polyurethane (PU) membrane matrix, has been developed for the selective quantification of monohydrogen phosphate ions. The best performing membrane contained PVC, PU, ionophore, and nitrophenyl octyl ether as a plasticizer in the ratio 32.2:2.6:65.1 (w/w, %). It exhibited a near-Nernstian slope of 31.0 ± 1.0 mV/decade of activity for HPO42− ions in the concentration range of 1.0 × 10−6 to 1.0 × 10−2 M at pH 7.4. The detection limit of the electrode was 8.4 × 10−7 M and the life time was six weeks. The electrode displayed excellent selectivity for monohydrogen phosphate over other anions and the selectivity sequence was determined as HPO42− > SO42− > Ac− > NO3− > ClO4− > Cl− > I−. The selective electrode for the monohydrogen phosphate ions was evaluated with a standard reference material (SRM 1548) and the titration of the sample solution. |
| |
Keywords: | Ionophore Monohydrogen phosphate All-solid-state-membrane electrode Potentiometric titration Ion-selective electrode Anion sensor |
本文献已被 ScienceDirect 等数据库收录! |
|