首页 | 本学科首页   官方微博 | 高级检索  
     


Two analyte calibrations from the transient response of a single potentiometric sensor employed with the SIA technique
Authors:Raul Cartas  Andrey Legin
Affiliation:a Sensors and Biosensors Group, Chemistry Department, Universitat Autònoma de Barcelona, Edifici Cn, 08193 Bellaterra, Barcelona, Spain
b Chemistry Dept, St. Petersburg University, Universitetskaya nab. 7/9, 199034 St. Petersburg, Russia
Abstract:Simultaneous quantification of Cd2+ and Pb2+ in solution has been correctly targeted using the kinetic information from a single non-specific potentiometric sensor. Dual quantification was accomplished from the complex information in the transient response of an electrode used in a Sequential Injection Analysis (SIA) system and recorded after step injection of sample. Data was firstly preprocessed with the Discrete Wavelet Transform (DWT) to extract significant features and then fed into an Artificial Neural Network (ANN) for building the calibration model. DWT stage was optimized regarding the wavelet function and decomposition level, while the ANN stage was optimized on its structure. To simultaneously corroborate the effectiveness of the approach, two different potentiometric sensors were used as study case, one using a glass selective to Cd2+ and another a PVC membrane selective to Pb2+.
Keywords:Wavelet transform   Artificial neural networks   Multi-analyte calibration   SIA   Potentiometry
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏