摘 要: | A smallbore chemically bonded alkyl stationary phase column of 2 mm I. D.(inner diameter) and 15 cm in length having an efficiency of 6400—7600 theoretical plates has been developed. The column officicncy is nearly the same as that of routine analytical column of 4.6—5.0 mm I. D. The amount of cluant used for the smallbore column, however, is less than about one fifth that for the ordinary columns.Using the reaction of monochlorosilane with 5 μm accumulated silica bead produced in Qingdao Ocean Chemical Factory instead of trichlorosilane and irregular silica gel, we have. prepared a nonpolar chemically bonded phase packing which has a higher carbon content and more obvious hydrophobicity. Moreover, the reproducibility of the preparation is better.The equipment and assembly necessary for getting high efficiency with the smallbore column are discussed briefly. Using Model SW-1 smallbore column HPLC instrument developed in the institute with the Model K-1 column fittings and injector, the detector with
|