首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Plasma-polymerized membrane electrode for the determination of dextromethorphan and dimemorfan
Authors:N Hazemoto  S Ishizaka  M Haga  Y Kato  S Kurosawa  N Kamo  Y Kobatake
Abstract:Ion-selective electrodes (ISEs) responsive to the antitussives dextromethorphan and dimemorfan were constructed by the fixation of an ion-exchanger, ammonium tetraphenylborate, on a Millipore membrane by means of a plasma-polymerization technique. The electrodes showed a Nernstian response over the range of 10(-5)-10(-2) M dextromethorphan and dimemorphan, and the working pH range was 5-7. The interference from common cations such as Na+, K+ and Ca2+ was negligible but some organic cations interfered weakly. The electrodes were applied successfully for the determination of the drugs in pharmaceutical preparations.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏