Label-free monitoring of site-specific DNA cleavage by EcoRI endonuclease using cyclic voltammetry and electrochemical impedance
Authors:Yan Jin
Institution:Key Laboratory of Analytical Chemistry for Life Science of Shaanxi Province, School of Chemistry and Materials Science, Shaanxi Normal University, Xi’an 710062, PR China
Abstract:Cyclic voltammetry (CV) combined with electrochemical impedance spectroscopy (EIS) were proposed to monitor the site-specific DNA cleavage by EcoRI endonuclease without using external label. The alteration of CV and EIS signal demonstrated that double-strands (dsDNA) contain recognition sequence was cleaved by EcoRI endonuclease. Real-time monitoring indicated that the dsDNA was cleaved by EcoRI more than 90% after 2 h of enzyme digestion time. Control experiment showed that the DNA cleavage by EcoRI endonuclease is site-specific for DNA sequence. Experimental results demonstrated that the efficiency of EcoRI cleavage was highly dependent on the concentration of EcoRI concentration in the range from 0.04 to 0.4 U μL−1 with one almost linear relationship.
Keywords:EcoRI endonuclease  Site-specific  DNA cleavage  Label-free  Electrochemical impedance  Cyclic voltammetry
本文献已被 ScienceDirect 等数据库收录!