首页 | 本学科首页   官方微博 | 高级检索  
     


Improved high-performance liquid chromatographic method for the simultaneous determination of caffeine and its N-demethylated metabolites in plasma using solid-phase extraction
Authors:R Hartley  I J Smith  J R Cookman
Abstract:An improved high-performance liquid chromatographic method for the simultaneous determination of caffeine and its N-demethylated metabolites in plasma is described. Excellent resolution of all components is provided by reversed-phase chromatography using a mobile phase consisting of 1% acetic acid-methanol (83:17) at a flow-rate of 2.7 ml/min, in conjunction with a Waters Assoc. Nova-Pak C18 column which was protected by a Waters Assoc. Guard-Pak precolumn module containing a Guard-Pak CN cartridge. Rapid extraction of caffeine and the dimethylxanthines from plasma was achieved using reversed-phase octadecylsilane bonded-silica columns (Bond-Elut C18). With only 100 microliters of sample, plasma levels in the region of 50 ng/ml for the dimethylxanthines and 100 ng/ml for caffeine can be determined using ultraviolet detection at 273 nm. The method has been used for measuring umbilical cord plasma samples to provide information regarding foetal exposure to caffeine and its metabolites and is also suitable for therapeutic drug monitoring of caffeine and theophylline levels in the treatment of neonatal apnoea.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏