首页 | 本学科首页   官方微博 | 高级检索  
     


Analysis of 2,4,6-trinitrotoluene, pentaerythritol tetranitrate and cyclo-1,3,5-trimethylene-2,4,6-trinitramine using negative corona discharge ion mobility spectrometry
Authors:Khayamian T  Tabrizchi M  Jafari M T
Affiliation:Department of Chemistry, Isfahan University of Technology, Isfahan 84154, Iran
Abstract:In this study, the capability of negative corona discharge ion mobility spectrometry (IMS) for quantitative magnitude of several explosives including 2,4,6-trinitrotoluene (TNT), pentaerythritol tetranitrate (PETN) and cyclo-1,3,5-trimethylene-2,4,6-trinitramine (RDX) has been evaluated for the first time. The total current obtained with the negative corona discharge was about 100 times larger than that of IMS based on 63Ni, which results in a lower detection limit and a wider linear dynamic range. The detection limits for PETN, TNT and RDX were 8×10−11, 7×10−11 and 3×10−10 g, respectively. The calibration plots for these explosives showed linear dynamic ranges of about four orders of magnitude.
Keywords:Negative corona discharge   Explosives   Ion mobility spectrometry
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号