首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A capillary electrophoresis detection scheme for underivatized amino acids based on luminol-BrO- chemiluminescence system
Authors:Yang Weiping  Zhang Zhujun  Deng Wei
Institution:College of Chemistry and Materials Science, Shaanxi Normal University, Xi'an 710062, China
Abstract:A novel chemiluminescence (CL) detection scheme has been developed for detecting underivatized amino acids following capillary electrophoresis (CE) separation. This detection was based on the inhibitory effect of amino acids on the CL reaction between luminol and BrO in alkaline aqueous solution. Detection of amino acids was accomplished with a borate-based background electrolyte at pH 9.2. The luminol was used as a component of the separation carrier electrolyte. Parameters affecting CE-CL separation and detection, such as the pH value, the concentration of electrolyte and CL reagent on the resolution were optimized. The relative standard deviation for the analysis of amino acids was less than 1.5% for the migration time and 4% for the peak height. The mass limits of detection were from 7 to 144 fmol for the 7 amino acids. This method has been applied of 7 amino acids in amino acid injection.
Keywords:Capillary electrophoresis  Chemiluminescence  Amino acids  BrO&minus  