首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Blood flow and oxygen transfer rate of an outside blood flow membrane oxygenator
Authors:Noriaki Matsuda  Kiyotaka Sakai  
Institution:

Chemical Engineering Department, Waseda University, 3-4-1 Okubo, Shinjuku-ku, Tokyo 169-8555, Japan

Abstract:We evaluated effects of the number of tied hollow fibers of an outside blood flow membrane oxygenator with cross-wound hollow fibers on the blood flow pattern and oxygen transfer rate. The number of tied hollow fibers in a bundle was varied from one to six, and the blood flow pattern was observed by X-ray computed tomography. The oxygen transfer rate and blood pressure drop were measured by in vitro experiments using bovine blood. Uniform blood flow patterns were obtained for each number of tied hollow fibers. A decrease in the number of tied hollow fibers caused more effective contact of blood with the tied hollow fibers and oxygen transfer rate was enhanced, demonstrating that single hollow fiber was the most effective. Empirical equations were obtained based on these results and optimum structure parameters of the membrane oxygenator were determined by simulation analysis. Optimum membrane surface area and axial jacket length of the oxygenator were 3.0 m2 and 320 mm, respectively, at a hollow fiber outside diameter of 250 μm.
Keywords:Artificial lung  Gas separation  Fiber membranes  Blood flow  Optimum design
本文献已被 ScienceDirect 等数据库收录!