Determination of halogenated mono-alcohols and diols in water by gas chromatography with electron-capture detection |
| |
Authors: | Matthew B M Anastasio C |
| |
Institution: | Department of Land, Air and Water Resources, University of California, Davis, 95616-8627, USA. |
| |
Abstract: | We have developed an analytical method for the detection of halogenated alcohols in water with particular focus on 3-chloro-1,2-propanediol and 3-bromo-1,2-propanediol. In this method the target analytes are extracted from water, derivatized with heptafluorobutyric anhydride, and then analyzed with gas chromatography with electron-capture detection. The effects of water, pH and seawater constituents on the method were investigated. Method detection limits for a 5 ml aqueous sample ranged from 0.14 microg l(-1) for 2-bromo-1,3-propanediol to 1.7 microg l(-1) for 1,3-dichloro-2-propanol (1,3DCP). |
| |