Advances in ion chromatography: speciation of μ 1−1 levels of metallo-cyanides using ion-interaction chromatography |
| |
Authors: | Paul R. Haddad Chatvalee Kalambaheti |
| |
Affiliation: | Department of Analytical Chemistry, University of New South Wales, P.O. Box 1, Kensington, New South Wales 2033 Australia |
| |
Abstract: | Fe(CN)4?6, Cu(CN)3?4, Co(CN)3?6, Fe(CN)3?6, Ni(CN)2?4 and Cr(CN)3?6 are determined by ion-interaction chromatography using a C18 column and methanol-tetrahydrofuran-10 mM phosphate buffer (pH 7.9) (25 + 1 + 74, v/v/v) containing 5 mM tetrabutylammonium hydroxide as mobile phase, with spectrophotometric detection at 214 nm. Detection limits are in the range 0.01–0.5 mg 1?1. In an alternative approach, an automated on-line sample preconcentration technique is used wherein a 2-ml volume of sample containing metallo-cyanides is loaded onto a C18 precolumn which has been equilibrated with the above mobile phase. The bound solutes are then eluted from the precolumn to a C18 analytical column where they are separated using the same mobile phase as employed to equilibrate the precolumn. Detection limits are in the rate 0.08–1.58 μg 1?1 and calibration graphs are linear up to 200 μg 1?1. The preconcentration step is shown to give quantitative recoveries for all species except Fe(CN)4?6 and (CN)3?4. The iron(II) complex does not bind quantitatively to the precolumn, and extensive studies with the copper complex suggested that low recoveries were due to dissociation and ligand-exchange reactions occurring during the chromatographic separation process. Negative interference effects were observed for Cl? and SO2?4 when present at a level of 250 mg 11?, and UV-absorbing anions such as Br?, SCN?, NO?2 and NO?3 caused positive interference when present at concentrations as low as 1 mg 1?1. The negative interferences could be reduced by diluting the sample and the positive interferences could be eliminated by incorporating an additional step in the preconcentration process, in which UV-absorbing anions bound to the precolumn after sample loading were eluted selectively using an eluent consisting of 10 mM NaCl in phosphate buffer (pH 6.7). |
| |
Keywords: | Chromatography Ion exchange Cyanides Ion-interaction chromatography Metallo-cyanides Preconcentration |
本文献已被 ScienceDirect 等数据库收录! |
|