首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Amperometric determination of glucose in undiluted food samples
Authors:A Amine  GJ Patriarche  G Marrazza  M Mascini
Institution:Free University of Brussels (ULB), Institute of Pharmacy, Campus Plaine, C.P. 205 / 6, Bd du Triomphe, B-1050 Brussels Belgium;Dipartimento Sanita Pubblica, Epidemiologia e Chimica Analitica Ambientale, Sezione Chimica Analitica, Via Gino Capponi 9, 50121 Firenze Italy
Abstract:Glucose oxidase is immobilized onto a cellulose acetate membrane by glutaraldehyde linkage, and the membrane is used to cover the platinum electrode of a hydrogen peroxide sensor. A silanized polycarbonate membrane then covers the enzyme layer, and extends the linear calibration range to higher concentrations. The sensor, when incorporated into a flow-injection system, allows the determination of glucose at levels up to 1 M in soft drinks at a rate of 60 samples h?1 without sample dilution.
Keywords:Amperometry  Flow system  Glucose  Foods  Soft drinks  Enzyme electrodes
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏