Flow-injection determination of d-mannitol with immobilized mannitol dehydrogenase |
| |
Authors: | Nobutoshi Kiba Yukari Inoue Motohisa Furusawa |
| |
Affiliation: | Department of Applied Chemistry and Biotechnology, Faculty of Engineering, Yamanashi University, Kofu 400 Japan |
| |
Abstract: | A flow-injection system is described for the determination of d-mannitol. Mannitol dehydrogenase is immobilized on poly(vinyl alcohol) beads and packed in a column (5 cm × 4 mm i.d.). The NADH formed is detected fluorimetrically. The response is linear between 5 × 10?7 and 1 × 10?4 M mannitol and the detection limit is 1 × 10?7 M. The throughput is 30 samples per hour. The reactor is stable for at least 8 weeks. |
| |
Keywords: | Flow system Fluorimetry Mannitol |
本文献已被 ScienceDirect 等数据库收录! |