Optimization of a rapid capillary electrophoresis ESI-IT tandem mass spectrometry method for the analysis of short-chain carnitines in human plasma
Authors:Claudia Desiderio  Antonella De Rossi  Rosanna Inzitari  Angelo Mancinelli  Diana Valeria Rossetti  Massimo Castagnola  Irene Messana
Institution:(1) Istituto di Chimica del Riconoscimento Molecolare-Sez di Roma, Consiglio Nazionale delle Ricerche, c/o Istituto di Biochimica e Biochimica Clinica, dell’Università Cattolica del Sacro Cuore, L.go F. Vito 1, 00168 Roma, Italy;(2) Istituto di Metodologie Chimiche, Consiglio Nazionale delle Ricerche, Area della Ricerca Roma 1, Via Salaria km 29.300, 00016 Monterotondo Stazione, Roma, Italy;(3) Istituto di Biochimica e Biochimica Clinica, Università Cattolica del Sacro Cuore, L.go F. Vito 1, 00168 Roma, Italy;(4) Dipartimento Scientifico SIGMA-TAU, Via Pontina km 30,400, 00040 Pomezia, Roma, Italy;(5) Dipartimento di Scienze Applicate ai Biosistemi, Università di Cagliari, Campus Monserrato, 09042 Monserrato, Cagliari, Italy
Abstract:A capillary electrophoresis (CE) method coupled to electrospray ionization ion trap tandem mass spectrometry (ESI-IT-MS/MS) is described for the rapid analysis of carnitine, acetylcarnitine, and propionylcarnitine in human plasma. Optimization of the procedure was achieved by a reduced sample pretreatment and after examining several physicochemical parameters that influence both the CE separation and the MS analytes detection. The analysis of total carnitine in human plasma after hydrolysis of short-chain metabolites is also shown. The analysis of carnitine and metabolites was obtained in less than 10 min using a 200 mM ammonium formate buffer, pH 2.5, with high sensitivity and specificity using the MS detection in product ion scan mode. The method was tested for quantitative recovery using dialyzed human plasma as matrix and showed linearity in the concentrations ranges 20–160, 1–32, and 0.25–8 μM for carnitine, acetylcarnitine, and propionylcarnitine with (squared) correlation coefficients of 0.9984, 0.9995, and 0.9991, respectively. The intraday and intermediate analysis repeatability and accuracy are within 15% of relative standard deviation (RSD) at low, medium, and high concentration and within/or slight exceeding 20% at the lower limit of quantitation (LLOQ). The method is sensitive for determining carnitine and its metabolites in human plasma with high specificity.
Keywords:Capillary electrophoresis  Mass spectrometry  Carnitines  Biological fluids  Sample preparation
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号