Feasibility of eliminating interferences in graphite furnace atomic absorption spectrometry using analyte transfer to the permanently modified graphite tube surface
Affiliation:1. Instituto de Quı́mica, Universidade Federal do Rio Grande do Sul, 91501-970 Porto Alegre, RS, Brazil;2. Departamento de Quı́mica, Universidade Federal de Santa Catarina, 88040-900 Florianópolis, SC, Brazil;1. Department of Chemistry, Kumamoto University, 2-39-1 Kurokami, Kumamoto 869-8555, Japan;2. Environmental Science Research Laboratory, Central Research Institute of Electric Power Industry, 1646 Abiko, Abiko 270-1194, Japan;1. College of Precision Instrument and Opto-Electronics Engineering, Tianjin University, Tianjin 300072, China;2. Key Laboratory of Opto-electronics Information Technology (Tianjin University) Ministry of Education, Tianjin 300072, China
Abstract:A procedure is proposed to avoid spectral and/or non-spectral interferences in graphite furnace atomic absorption spectrometry (GF AAS) by transferring the analyte during the pyrolysis stage from a solid sampling platform to the graphite tube wall that has been coated with a permanent modifier, e.g. by electrodeposition of a platinum-group metal. The direct determination of mercury in solid coal samples was chosen as a model to investigate the feasibility of this idea. The graphite tube surface was coated with palladium and the analyte was transferred from the solid sampling platform to the tube wall at a temperature of 500±50 °C. A characteristic mass of m0=64 pg Hg was obtained for an atomization temperature of 1300 °C, proposing a quantitative transfer of the analyte to the tube wall. Calibration against aqueous mercury standards was not feasible as this element was lost in part already during the drying stage and could not be trapped quantitatively on the modified graphite tube surface. However, the results for all except one of the coal reference materials were within the 95% confidence interval of the certificate when the slope of a correlation curve between the integrated absorbance, normalized for 1 mg of sample, and the certified value for mercury was used for calibration. A detection limit of 0.025–0.05 μg g−1 Hg in coal, calculated from three times the standard deviation of the investigated coal samples, could be obtained with the proposed method. The spectral interference due to excessive background absorption in the direct determination of mercury in coal could be eliminated completely. It is expected that this analyte transfer can be used in a similar way to eliminate other spectral and/or non-spectral interferences in the GF AAS determination of other volatile analytes.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号