Calculation of attenuation and x‐ray fluorescence intensities for non‐parallel x‐ray beams
Abstract:With the nowadays widespreaded use of x‐ray optics in x‐ray fluorescence analysis, large convergence or divergence angles can occur. This experimental situation violates a basic assumption of the usual fundamental parameter quantification procedure. In order to take beam divergences in micro x‐ray fluorescence analysis into account, a way of calculating fluorescence intensities numerically by Monte Carlo integration is described. For three examples of typical micro‐XRF set‐ups the fluorescence intensities and their deviation from the parallel beam geometry are calculated. Furthermore, we propose a new approach with ‘equivalent angles’ which correct for the beam divergences in fundamental parameter methods. Copyright © 2003 John Wiley & Sons, Ltd.