首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Degradation of plasticized PVC for biomedical disposable device under soft x‐ray irradiation
Abstract:Plasticized poly(vinyl chloride) (PVC) used for biomedical disposable devices was studied in the non‐sterilized state after different exposure times to soft x‐ray irradiation in a commercial photoelectron spectrometer (Al Kα, 15 kV/300 W; Mg Kα, 12 kV/240 W) by XPS surface analysis. The detailed spectra of C 1s, Cl 2p and O 1s have been recorded and processed. Irradiation with soft x‐rays induces a clear decrease of the total Cl 2p intensity, an increase of total C 1s intensity and a doubling of the O 1s intensity after 45 min of irradiation with Al Kα (300 W). Irradiation with Mg Kα (240 W) is slightly less damaging. These results can be interpreted with the classical PVC degradation model, e.g. bond cleavage with the formation of HCl gas, although the Cl 2p high‐resolution spectra reveal the formation of an additional side‐product, probably CaCl2. For further studies of plasticized PVC using XPS surface analysis it can be concluded that a complete analysis of a polymer sample should not take >10 min of x‐ray exposure in order to avoid notable polymer degradation. Copyright © 2003 John Wiley & Sons, Ltd.
Keywords:XPS  medical disposable device  PVC  soft irradiation  degradation
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号