首页 | 本学科首页   官方微博 | 高级检索  
     


System for determination of ultrasonic wave speeds and their temperature dependence in liquids and in vitro tissues
Authors:Yost William T  Macias Brandon R  Cao Peihong  Hargens Alan R  Ueno Toshiaki
Affiliation:NASA-Langley Research Center, Hampton, Virginia 23681-2199, USA.
Abstract:An interferometric technique capable of accurately measuring wave speed in liquids is reported. The hardware is adapted from a design to measure nonlinear responses of biological tissues to pressure changes (pressure derivatives) and temperature changes (temperature derivatives). It is used with the highly sensitive variable frequency pulsed phase-locked loop (VFPPLL) instrument. The system uses well-understood and well-characterized components and systems. The apparatus covers a temperature range from below 5 degrees C to above 45 degrees C. The system with the high-sensitivity VFPPLL is capable of measurement of wave speed to an uncertainty of less than 0.1%, and changes in wave speed to better than 0.001%. The transducer is an undamped temperature-characterized PZT-5A 500-kHz plate, whose output is corrected for off-resonance operation and for diffraction effects. To test the accuracy of the technique, measurement of ultrasonic compressional wave speed in water at temperatures from 10 degrees C to 45 degrees C are reported, with an estimated uncertainty of 0.07% and a temperature uncertainty of 0.15 degrees C. The agreement between mean values and literature values is better than 0.05%.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏