Energy calibration in electron spectroscopy and the re-determination of some reference electron binding energies |
| |
Authors: | R.J. Bird P. Swift |
| |
Affiliation: | Shell Research Limited, Thornton Research Centre, P.O. Box 1, Chester Gt. Britain |
| |
Abstract: | Measured binding energies were found to vary with the choice of reference binding energies used in instrument calibration. Inadequate linearity of the spectrometer energy scale was eliminated as a source of the error. Primary calibration methods were considered, and tried. The instrument was finally set up by the two X-ray wavelength method, using Al Kα and the Ag Lα,β radiations. Subsequent re-determination of electron binding energies for noble metals and copper showed differences from previously adopted values that would be important in calibration work. |
| |