Determination of nanomolar levels of guanine by differential-pulse adsorptive cathodic stripping voltammetry of its copper(II) complex |
| |
Authors: | R M Shubietah A Z Abu Zuhri and A G Fogg |
| |
Institution: | (1) Chemistry Department, Faculty of Science, UAE University, P.O. Box 17551, Al-Ain, United Arab Emirates;(2) Chemistry Department, Loughborough University of Technology, LE 11 3TU Loughborough, Leicestershire, UK |
| |
Abstract: | Guanine is determined at the 5.0×10–10 –2.0×10–7 mol/l level by differential-pulse adsorptive stripping voltammetry at a hanging mercury drop electrode using the reduction peak of its copper (II) complex at –0.21 V vs. Ag-AgCl electrode. The optimum analytical conditions were found to be Britton-Robinson buffer solution (pH 4.8), an accumulation potential of 0.0 V and an accumulation time of 3 min. Under these conditions, the detection limit is 5.0×10–10 mol/l and the relative standard deviation 2.6% for 1.0×10–7 mol/l guanine. The method is compared with the previous voltammetric methods. The presence of some purine derivatives does not interfere. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|