首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Chemiluminescence Sensor for the Determination of Hydroxylamine by Electrostatically Immobilizing Luminol and Periodate
Authors:ZHANG Si-chun  Zhou Guo-jun  JU Huang-Xian
Institution:Department of Chemistry, Institute of Analytical Science, State Key Laboratory of Coordination Chemistry, Nanjing University, Nanjing 210093, P. R. China
Abstract:A novel chemiluminescence(CL) sensor, which can be used for hydroxylamine determination in combination with flow injection analysis, was developed by electrostatically immobilizing luminol and periodate on anion-exchange resin respectively. Hydroxylamine was sensed by its enhancing effect on the weak CL reaction between luminol and periodate, which were eluted from the ion exchange column. The response of the sensor to hydroxylamine was linear in the concentration range of 8.0×10-8-2.0×10-6 mol/L with a detection limit of 4.0×10-8 mol/L hydroxylamine(3σ). The relative standard deviation(RSD) was 2.0% for 9 repetitive determinations at a hydroxylamine concentration of 5.0×10-7 mol/L. The sensor could be reused for over 400 times with a good reproducibility and was used to determine hydroxylamine in wastewater.
Keywords:Chemical sensor  Chemiluminescence  Luminol  Periodate  Hydroxylamine
本文献已被 CNKI 万方数据 等数据库收录!
点击此处可从《高等学校化学研究》浏览原始摘要信息
点击此处可从《高等学校化学研究》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏