The response of halide ion-selective electrodes to redox systems : A comparison between silver/silver halide electrodes and silver sulphide- based halide ion-selective electrodes |
| |
Authors: | C. Harzdorf |
| |
Affiliation: | Bayer AG, Anorganisch analytisches Laboratorium, D-5090 LeverkusenW. Germany |
| |
Abstract: | Silver/silver chloride and bromide electrodes, prepared by anodizing ordinary silver electrodes, and the corresponding ion-selective electrodes based on silver sulphide, were tested for their susceptibility towards redox systems. It proved that the latter type of electrode responded significantly to strong oxidants. In contrast, the silver/silver halide types were highly resistant to redox interference provided that the silver halide layer was free from open pores. This could be achieved by generation of sufficiently thick layers and by selection of suitable current densities during electrodeposition (<20 mA cm-2). The interrelation between the conditions of silver chloride film generation and redox resistance of the resulting electrodes is described in detail. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |
|