首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of creatinine in clinical samples with a creatininase reactor and an ammonia probe
Authors:Marco Mascini  Giuseppe Palleschi
Institution:Istituto Chimica Analitica, Università di RomaItaly
Abstract:A commercial nylon coil with immobilized creatininase in conjunction with a potentiometric ammonia probe is used in a continuous flow apparatus for the determination of creatinine in blood plasma and urine. The analytical characteristics of the coil are evaluated and its specific enzymatic activity is calculated. The determination of creatinine in plasma is viable provided that samples are processed quickly after they have been taken and that plasma ammonia is evaluated just before the creatinine determination. Analyses of urine require prior separation of ammonia. For plasma, 40 samples/hour can be processed with confidence. The precision of the procedure depends strongly on the plasma ammonia level and is about 5% for normal conditions.
Keywords:
本文献已被 ScienceDirect 等数据库收录!