首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A flow injection procedure based on solenoid micro-pumps for spectrophotometric determination of free glycerol in biodiesel
Authors:Silva Sidnei G  Rocha Fábio R P
Institution:a Instituto de Química, Universidade de São Paulo, PO Box 26077, 05513-970, São Paulo, SP, Brazil
b Centro de Energia Nuclear na Agricultura, Universidade de São Paulo, PO Box 96, 13400-970, Piracicaba, SP, Brazil
Abstract:A flow system designed with solenoid micro-pumps is proposed for fast and greener spectrophotometric determination of free glycerol in biodiesel. Glycerol was extracted from samples without using organic solvents. The determination involves glycerol oxidation by periodate, yielding formaldehyde followed by formation of the colored (3,5-diacetil-1,4-dihidrolutidine) product upon reaction with acetylacetone. The coefficient of variation, sampling rate and detection limit were estimated as 1.5% (20.0 mg L−1 glycerol, n = 10), 34 h−1, and 1.0 mg L−1 (99.7% confidence level), respectively. A linear response was observed from 5 to 50 mg L−1, with reagent consumption estimated as 345 μg of KIO4 and 15 mg of acetylacetone per determination. The procedure was successfully applied to the analysis of biodiesel samples and the results agreed with the batch reference method at the 95% confidence level.
Keywords:Flow analysis  Multicommutation  Solenoid micro-pumps  Biodiesel  Glycerol
本文献已被 ScienceDirect PubMed 等数据库收录!