Determination of beer aroma compounds using headspace solid-phase microcolumn extraction |
| |
Authors: | Hrivňák Ján Smogrovičová Daniela Nádaský Pavol Lakatošová Jana |
| |
Affiliation: | a The Plant Production Research Center Pieš?any, Institute of Viticulture and Enology, Matúškova 25, 831 01 Bratislava, Slovak Republic b Faculty of Chemical and Food Technology, Slovak University of Technology, Radlinského 9, 812 37 Bratislava, Slovak Republic |
| |
Abstract: | A rapid sampling technique for the analysis of beer aroma compounds is described. The headspace (10 ml) is passed through the microcolumn filled with 5 mg of Tenax TA and thermally desorbed in a modified GC inlet (modification is described). Eight compounds (from acetaldehyde to 2-phenylethanol) in four beer samples were analyzed. The correlation coefficients (r2), repeatability (RSD) and limits of detection (LOD) were 0.9973-0.9994, 2.1-6.9% and 0.00002-0.13 mg/l, respectively. The methodology can be useful for routine beer sample analysis. |
| |
Keywords: | Gas chromatography Solid-phase microcolumn extraction Beer Volatile organic compounds Adsorption Tenax |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|