首页 | 本学科首页   官方微博 | 高级检索  
     


Selective and non-selective determination of heavy metal ions in flowing solutions by fast stripping cyclic voltammetry.
Authors:Parviz Norouzi  Mansoor Namazian  Alireza Badiei
Affiliation:Department of Chemistry, Faculty of Science, University of Tehran, Tehran, Iran. Norouzi@khayam.ut.ac.ir
Abstract:A simple and fast stripping voltammetric detection method has been designed for selective and non-selective measurements of heavy metal ions in a flow-injection system. A special computer numerical method is introduced for calculating the analyte signal and noise reduction, where the signal is calculated based on the partial and total charge exchange at electrode surface. For a selective determination, the currents are integrated in the range of the oxidation and reduction of the analyte. For non-selective measurements, the integration range is set for the whole potential scan range (including oxidation and reduction of the Au surface). The time for stripping has been shown to be less than 300 ms. The main advantages of the detection method are as follows: first, removal of oxygen from the measured solution is not required; second, it is sufficiently fast for the determination of heavy metal ions in various chromatographic analysis methods. The limit of detection for tested ions was between 3 x 10(-9) and 6 x 10(-10) M, and the relative standard deviation at 50 ppb Pb2+ was 4.7% for 10 runs.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号