首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Mass spectrometric analysis of high-purity carbon dioxide
Authors:Parkinson R T  Wilson R E
Institution:Master General of Ordnance Inspectorates, Chemical Inspectorate, Royal Arsenal Woolwich, London S.E.18, U.K.
Abstract:A high-vacuum, low-temperature, continuous separation technique has been used in conjunction with a mass spectrometer for the analysis of carbon dioxide containing vpm amounts of H(2), He, CH(4), Ne, N(2), CO, O(2) and Ar. The method relies on the condensation of carbon dioxide on the walls of a glass U-tube, cooled in liquid nitrogen, connected between an inlet and the ion source. A high-pressure carbon dioxide sample thus enters the inlet leak but only the impurities pass through the U-tube and reach the ion source, resulting in considerable gain in sensitivity and elimination of interference from carbon dioxide. The sensitivity of the method is several orders of magnitude better than the normal mass spectrometric method.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号