Determination of Cr(VI) by portable tungsten coil electrothermal atomic absorption spectrometer after surfactant-assisted dispersive liquid-liquid microextraction |
| |
Authors: | Xiaodong Wen Shengchun Yang Haizhu Zhang Xia Zhao Jie Guo |
| |
Affiliation: | 1. College of Pharmacy and Chemistry, Dali University, Dali, Yunnan 671000, Chinawenxdong129@163.com;3. College of Pharmacy and Chemistry, Dali University, Dali, Yunnan 671000, China |
| |
Abstract: | As a new developed instrument, a portable tungsten coil electrothermal atomic absorption spectrometer (W-coil ET-AAS) was first coupled with surfactant assisted dispersive liquid–liquid microextraction (SA-DLLME) to improve its analytical performance and expand its applications in this work. SA-DLLME was very simple, rapid and the extraction efficiency was considerably improved by the effect of surfactant, which was suitable to be coupled with the portable instrument in field analysis. After SA-DLLME, concentrated chromium in organic phase was directly determined on W-coil atomiser. The influence factors relevant to SA-DLLME and instrumental conditions were studied systematically. Under the optimal conditions, the limit of detection (LOD) for Cr(VI) was 0.016 µg L?1, with sensitivity enhancement factor (EF) of 107. The relative standard deviation (RSD) for seven replicate measurements of 0.5 µg L?1 of Cr(VI) was 4.6%. The recoveries for the spiked samples were in the acceptable range of 96.8–104%. The rapid, simple and high effective method greatly improved the sensitivity of this portable spectrometer for the determination of Cr(VI) and was applied to the analysis of ultra-trace Cr(VI) in real and certified water samples with satisfactory results. |
| |
Keywords: | portable tungsten coil electrothermal atomic absorption spectrometer preconcentration Cr(VI) surfactant-assisted dispersive liquid–liquid microextraction |
|
|