Sensitive and Selective Fluorescence Detection of Phenoxyacid Herbicides by Liquid Chromatography with On-Line Ion-Pair Extraction |
| |
Authors: | C. De Ruiter W. A. Minnaard H. Lingeman E. M. Kirk U. A. Th. Brinkman R. R. Otten |
| |
Affiliation: | 1. Department of Analytical Chemistry , Free University , De Boelelaan 1083, 1081, HV Amsterdam, The Netherlands;2. Environmental Research Laboratory City of Amsterdam , Amstelveenseweg 88, 1075 XJ, Amsterdam, The Netherlands |
| |
Abstract: | Abstract Fluorescent probes, especially a newly synthesized N-substituted 1-cyanobenz[f]-isoindole quaternary ammonium fluorophore, were used as counter ions in a reaction detector for on-line ion-pair extraction of phenoxyacid herbicides. The probe was used in an on-line post-column set-up coupled to a reversed-phase chromatographic system. After separation on an C-8-bonded silica column using an aqueous methanol (pH 2.5) mobile phase, the herbicides were on-line deprotonated by post-column addition of a 10mM sodium phosphate buffer (pH 8.0), in which the probe was dissolved. Subsequently, the ion-pairs were extracted on-line with chloroform-1-butanol (80:20, v/v) and were monitored by fluorescence detection. Using this system, at least seven herbicides could be separated. The detection limits of 2,4-dichlorophe xyacetic acid and 2,4,5-trichlorophenoxyacetic acid were 400 pg (S/N = 3). The repeatability, based on peak height measurements, for 100ng injections was about 0.5%. Calibration curves were linear over the investigated range of 1–100ng, with correlation coefficients of 0.999 for the two analytes. Application to a drinking water sample is presented. |
| |
Keywords: | Phenoxyacid herbicides on-line ion-pair extraction water analysis fluorescence detection. |
|
|