首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Rapid determination of bromide in seawater samples by capillary ion chromatography using monolithic silica columns modified with cetyltrimethylammonium ion
Authors:Suzuki Atsushi  Lim Lee Wah  Hiroi Toshiya  Takeuchi Toyohide
Institution:Department of Chemistry, Faculty of Engineering, Gifu University, 1-1 Yanagido, Gifu 501-1193, Japan
Abstract:Monolithic silica capillary columns dynamically modified with quaternary ammonium ions were evaluated for the determination of bromide in seawater samples. A quaternary ammonium ion such as cetyltrimethylammonium ion was dynamically introduced onto monolithic silica surfaces. The first layer of the modifier was introduced by electrostatic interaction, whereas the second layer was introduced by hydrophobic interaction. The latter layer worked as the anion-exchange sites. The modified monolithic silica capillary columns could be used for rapid separation of inorganic anions. Separation of authentic mixture of five anions was achieved within a few minutes. The addition of small amount of the modifier in the eluent improved the repeatability of the retention time. Seawater samples could be directly injected onto the prepared capillary columns, and bromide could be determined to be 63 mg/L.
Keywords:Ion chromatography  Monolithic silica capillary columns  Cetyltrimethylammonium ion  Bromide  Seawater
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏