首页 | 本学科首页   官方微博 | 高级检索  
     


Microchip flow-injection analysis of trace 2,4,6-trinitrotoluene (TNT) using mercury-amalgam electrochemical detector
Authors:Wang Joseph  Pumera Martin
Affiliation:a Departments of Chemical and Materials Engineering and Chemistry and Biochemistry, Arizona State University, Tempe, AZ 85287, USA
b Department of Chemistry, Universitat Autonoma de Barcelona, 08193 Bellaterra, Barcelona, Spain
Abstract:This paper reports on a microfluidic device for the flow-injection/electrochemistry analysis of nitroaromatic explosive. The response is very fast (150 assays/h), highly sensitive (detection limit 7.0 μg L−1), reproducible and stable (R.S.D. = 2.0%; n = 30) and linear (over 20-100 μg L−1 range). Relevant experimental parameters have been optimized. The new microsystem offers great promise for on-site monitoring of TNT, with significant advantages of speed/warning, sample size, efficiency and cost. Most favorable S/N characteristics were obtained at the Hg/Au-amalgam end-channel detector.
Keywords:Explosives   Microfluidics   Flow-injection analysis   Electrochemistry
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏