首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of metal traces in tantalum by proton activation analysis
Authors:I O Konstantinov  B V Zatolokin  N N Krasnov  Yu G Sevastyanov  L S Volkova
Institution:(1) Institute of Physics and Power Engineering, Obninsk, (USSR)
Abstract:Procedures for determining traces of calcium, titanium, vanadium, chromium, iron, copper, niobium, molybdenum and tungsten in tantalum are described utilizing 11 MeV proton activation. The instrumental method of analysis with the use of a Ge(Li) detector and the radiochemical separation of the corresponding fractions are described. The formula for the calculation of the concentration of traces with the use of thick target yields is given. Experimental data on the thick target yields of radioisotopes44Sc,52Mn,56Co,65Zn,93mMo,96Tc are presented. Based on the experimental data, the detection limits are estimated at 1–10 ppb for the above mentioned metal traces in tantalum. The successful solution of the task of the simultaneous determination of nine elements in tantalum confirms the wide applicability of proton activation analysis.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏